Abstract

This report summarises the presentations and activities of the SELECTBIO Workshop on Rigor and Reproducibility in EV Research and Single EV Analysis held in San Diego, USA, in December 2021. The motivation for the session was the recognition that progress in the extracellular vesicle (EV) field is limited by the availability of rigorous and reproducible EV measurement tools. These tools are absolutely required for EVs to evolve from a research lab curiosity to something that will improve our ability to understand, diagnose, treat, and prevent disease. The program focused on guidelines for EV measurement and characterization as laid out in the recent MISEV2018 and MIFlowCyt-EV publications, their implementation in routine practice, and their continued evolution as new EV measurement technologies are introduced. The conclusion of the workshop was that more effort focused on pre-analytical issues and benchmarking of isolation methods is needed to strengthen collaborations and advance more effective biomarkers.

Keywords

Extracellular vesicle, exosomes, flow cytometry, calibration, standardization, reproducibility

The recent SELECTBIO Conference in San Diego (December 12-14, 2021) included a Workshop on Rigor and Reproducibility in EV Research and Single EV Analysis as part of the Extracellular Vesicles: Technologies and Investigations track. The motivation for the session was the recognition that progress in the extracellular vesicle (EV) field is limited by the availability of rigorous and reproducible EV measurement tools. These tools are absolutely required for EVs to evolve from a research lab curiosity to something that will improve our ability to understand, diagnose, treat, and prevent disease. The program focused on guidelines for EV measurement and characterization as laid out in the recent MISEV2018[1] and MIFlowCyt-EV[2] publications, their implementation in routine practice, and their continued evolution as new EV measurement technologies are introduced.

The starting point for discussion was a review of the MISEV2018 reporting guidelines and the MIFlowCyt-EV reporting framework for single EV flow cytometry. Both of these reporting guidelines are intended as recommendations to increase rigor and reproducibility in the EV field, and also act as a tool for editors and reviewers to assess the strengths and weaknesses of manuscripts, proposals, funding applications, and conference presentations. The MISEV guidelines cover, at a high level, all aspects of EV research, from biofluid collection through EV and EV cargo characterization to EV functional analysis[1]. MIFlowCyt-EV addresses the reporting of essential details of single-EV flow cytometry analysis required for data interpretation and reproducibility, and is an extension of the general MIFlowCyt reporting framework[3,4]. Key elements of these reporting guidelines include: (1) the use of a calibrated instrument so that measurements can be reported in standardized and biologically meaningful rather than arbitrary units; (2) inclusions of several essential controls to demonstrate the specificity of EV detection and cargo measurement; and (3) comprehensive reporting of methods and results, including data sharing, to enable others to reproduce and extend the work. Workshop attendees were introduced to the significant amount of educational material on these topics available on the website of the EV Flow Cytometry Working Group (https://www.evflowcytometry.org/), a group of EV researchers affiliated with ISEV, ISAC, and ISTH that developed and promoted these guidelines[5,6].

The discussion also covered challenges and opportunities for future advances in single-EV analysis. Rigorous and reproducible measurements of EV number, size and cargo by flow cytometry are possible[2,7-9] and routine with commercial assay kits[10-14], but instrument performance generally limits assay sensitivity. Most commercial flow cytometers have been designed to measure lymphocytes, but lab-built flow cytometers with single-molecule sensitivity have been possible for decades[15-17], and may soon be commercially available. The general lessons for assay development and validation using these single-molecule sensitive instruments still apply, including the need to fully report methodological details, demonstrate the specificity of EV and cargo detection through the appropriate controls, and report data in the appropriate calibrated units. This last point generated additional discussion centered around how to reconcile “top-down” calibration approaches developed for the analysis of cells with the “bottom-up” approach enabled by single-molecule detection.

Flow cytometer fluorescence calibration was initially developed for cell analysis several decades ago, with standards, protocols and software to allow users to readily report particle brightness in standard units (MESF, mean equivalent soluble fluorochromes; ERF, equivalent reference fluorochromes; or ABC, antibodies bound per cell)[18-21]. Approaches for fluorescence calibration are robust for the measurement of bright particles such as cells, and have been shown to have utility in standardization when extrapolated to the measurement of dim particles such as EVs on suitably sensitive instruments. Along with fluorescence calibration, the development of light scatter calibration methods has shown utility for the standardization and characterization of small particles[7,8,17,22,23]. When detecting light from very small numbers of photons (e.g., ~100 fluorochrome molecules), additional factors arise and can affect the uncertainty of these calibrations[7]. One of these is the natural effect of counting small numbers of anything (including photons, molecules, or particles), where Poisson-type counting statistics introduce a known but stochastic uncertainty into the results. Another has to do with the ways different instruments handle background. Sensitive detection of any analyte is most often limited by background, and definitions of the lower limits of detection (LoD) often reference the variation in the background signal (e.g., LOD is the mean of the background +3 standard deviations), whether molecules or particles are being detected. Most commercial instruments “subtract” the constant background measured by the instrument, but they often perform this with different methods, and cannot remove the variation (or “noise”) in the background, which generally remains. Moreover, beads used for calibration often have their own intrinsic background autofluorescence, the measurement of which will depend on the sensitivity of the instrument. Thus, while the “top-down” calibration approaches developed for cell analysis can be extended into the realm of very dim EVs and the instrument limits of detection, care is required to understand the uncertainties and avoid over-interpretation.

By contrast, single-molecule detection starts with the detection of a single molecule, and a particle that is brighter than a single molecule should be able to be reported as having some number of single molecules (the “bottom-up” approach)[24]. In practice, for single molecule-based detection approaches, issues of Poisson counting statistics and definition of background also dominate data interpretation, and pose challenges to report values calibrated in biologically meaningful units (e.g., molecules, antibodies, size). First, an inspection of the fluorescence intensity distribution of many detected single molecules generally shows a Log-normal distribution[25], which demonstrates that the photons detected to form that distribution are produced and detected stochastically, and that there are additional variabilities imposed by the measurement process. This means that intensity distributions of particles bearing one, two, three, or more fluorescent molecules can have significant overlap. Second, often the very sensitive detectors used for single-molecule detection have a limited dynamic range (the difference between the upper and lower LoD), and efforts to measure real biological samples, in which an individual EV may bear from zero to several hundred antibody molecules, can confront a limitation to measure brighter particles. Finally, while mathematical models can deconvolve such distributions post hoc to estimate the numbers of particles bearing zero, one, and two molecules (for example) or account for the effect of detector saturation for brighter particles, the ability to measure, rather than just detect, the estimated number of single molecules on a particle is subject to a number of assumptions and uncertainties.

From this perspective, the convergence of conventional flow cytometry technologies and the “top-down” measurement approach with high sensitivity, single molecule-based detection and the “bottom-up” approaches to measuring the low numbers of cargo molecules on many individual EVs is an exciting challenge in bioanalytical chemistry. As instrument sensitivities improve to strive for single molecule sensitivity, new challenges in high-throughput purification methods may need to be considered in order to remove all unbound label. With existing technologies, titration and dilution are usually sufficient to achieve this due to insensitivity to single molecules. Moreover, as other technologies are applied to measure single EVs, they will be challenged by many of these same issues, including (1) the development of appropriate concentration, size, and molecular cargo standards for single EV measurements; (2) the use of such standards to calibrate instrument response in absolute, rather than relative units, (3) the development of assays that report EV number, size, and molecular cargo in absolute units that can be compared across instruments and between labs; and (4) the sharing of instrument and assay methods and data in a manner that supports replication. These challenges must be solved to develop methods for EV measurement with the analytical rigor needed to understand their roles in health and disease, and to translate that understanding into useful new clinical diagnostics and therapeutic approaches.

DECLARATIONS

Author’s contributions

Conceptualization, writing: Nolan JP, Chiu DT, Welsh JA

Read and approved the manuscript: Nolan JP, Chiu DT, Welsh JA

Availability of data and materials

Not applicable.

Financial support and sponsorship

Work in the authors’ labs is funded by NIH 4UH3 CA241687, R01 GM140137, R44 DA044616, R44 GM136165, 4UH3 TR002874. JAW was supported by Intramural Research Program of the National Institutes of Health (NIH), National Cancer Institute, and Center for Cancer Research, NIH ZIA BC011502, NIH ZIA BC011503, and NIH 4UH3TR002881. JAW is an International Society for Advancement of Cytometry (ISAC) Marylou Ingram Scholar 2019-2023.

Conflict of interests

JPN holds equity in Cellarcus Biosciences Inc. DTC has financial interest and is a scientific founder and/or board member of the following companies and their respective affiliates: Micareo, Inc, Lamprogen, Inc, Cellectricon AB, and Fluicell AB.

Ethical approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Copyright

© The Author(s) 2022.

REFERENCES

1. Théry C, Witwer KW, Aikawa E, et al. Minimal information for studies of extracellular vesicles 2018 (MISEV2018): a position statement of the International Society for Extracellular Vesicles and update of the MISEV2014 guidelines. J Extracell Vesicles 2018;7:1535750.

2. Welsh JA, Van Der Pol E, Arkesteijn GJA, et al. MIFlowCyt-EV: a framework for standardized reporting of extracellular vesicle flow cytometry experiments. J Extracell Vesicles 2020;9:1713526.

3. Spidlen J, Breuer K, Brinkman R. Preparing a Minimum Information About a Flow Cytometry Experiment (MIFlowCyt) compliant manuscript using the International Society for Advancement of Cytometry (ISAC) FCS file repository (FlowRepository.org). Curr Protoc Cytom 2012;Chapter 10:Unit 10.18.

4. Lee JA, Spidlen J, Boyce K, et al. International Society for Advancement of Cytometry Data Standards Task Force. MIFlowCyt: the minimum information about a Flow Cytometry Experiment. Cytometry A 2008;73:926-30.

5. Welsh JA, Tang VA, van der Pol E, Görgens A. MIFlowCyt-EV: the next chapter in the reporting and reliability of single extracellular vesicle flow cytometry experiments. Cytometry A 2021;99:365-8.

6. der Pol E, Welsh JA, Nieuwland R. Minimum information to report about a flow cytometry experiment on extracellular vesicles: communication from the ISTH SSC subcommittee on vascular biology. J Thromb Haemost 2022;20:245-51.

7. Welsh JA, Jones JC, Tang VA. Fluorescence and light scatter calibration allow comparisons of small particle data in standard units across different flow cytometry platforms and detector settings. Cytometry A 2020;97:592-601.

8. van der Pol E, de Rond L, Coumans FAW, et al. Absolute sizing and label-free identification of extracellular vesicles by flow cytometry. Nanomedicine 2018;14:801-10.

9. Tian Y, Gong M, Hu Y, et al. Quality and efficiency assessment of six extracellular vesicle isolation methods by nano-flow cytometry. J Extracell Vesicles 2020;9:1697028.

10. Crooks ET, Almanza F, D'Addabbo A, et al. Engineering well-expressed, V2-immunofocusing HIV-1 envelope glycoprotein membrane trimers for use in heterologous prime-boost vaccine regimens. PLoS Pathog 2021;17:e1009807.

11. Oh CK, Dolatabadi N, Cieplak P, et al. S-Nitrosylation of p62 inhibits autophagic flux to promote α-synuclein secretion and spread in parkinson’s disease and lewy body dementia. J Neurosci 2022;42:3011-24.

12. Sandau US, Duggan E, Shi X, et al. Methamphetamine use alters human plasma extracellular vesicles and their microRNA cargo: an exploratory study. J Extracell Vesicles 2020;10:e12028.

13. Shpigelman J, Lao FS, Yao S, et al. Generation and application of a reporter cell line for the quantitative screen of extracellular vesicle release. Front Pharmacol 2021;12:668609.

14. Yan W, Cao M, Ruan X, et al. Cancer-cell-secreted miR-122 suppresses O-GlcNAcylation to promote skeletal muscle proteolysis. Nat Cell Biol 2022;24:793-804.

15. Nguyen DC, Keller RA, Jett JH, Martin JC. Detection of single molecules of phycoerythrin in hydrodynamically focused flows by laser-induced fluorescence. Anal Chem 1987;59:2158-61.

16. Soper SA, Shera EB, Martin JC, et al. Single-molecule detection of Rhodamine 6G in ethanolic solutions using continuous wave laser excitation. Anal Chem 1991;63:432-7.

17. Zhu S, Ma L, Wang S, et al. Light-scattering detection below the level of single fluorescent molecules for high-resolution characterization of functional nanoparticles. ACS Nano 2014;8:10998-1006.

18. Gaigalas AK, Wang L, Schwartz A, Marti GE, Vogt RF Jr. Quantitating fluorescence intensity from fluorophore: assignment of MESF values. J Res Natl Inst Stand Technol 2005;110:101-14.

19. Schwartz A, Wang L, Early E, et al. Quantitating fluorescence intensity from fluorophore: the definition of MESF assignment. J Res Natl Inst Stand Technol 2002;107:83-91.

20. Wang L, Hoffman RA. Standardization, calibration, and control in flow cytometry. Curr Protoc Cytom 2017;79:1.3.1-1.3.27.

21. Hoffman RA. Standardization, calibration, and control in flow cytometry. Curr Protoc Cytom 2005;Chapter 1:Unit 1.3.

22. der Pol E, van Gemert MJ, Sturk A, Nieuwland R, van Leeuwen TG. Single vs. swarm detection of microparticles and exosomes by flow cytometry. J Thromb Haemost 2012;10:919-30.

23. Rond L, Coumans FAW, Welsh JA, Nieuwland R, van Leeuwen TG, van der Pol E. Quantification of light scattering detection efficiency and background in flow cytometry. Cytometry A 2021;99:671-9.

24. Keller RA, Ambrose WP, Goodwin PM, Jett JH, Martin JC, Wu M. Single-molecule fluorescence analysis in solution. Appl Spectrosc 1996;50:12A-32A.

25. Mutch SA, Fujimoto BS, Kuyper CL, et al. Deconvolving single-molecule intensity distributions for quantitative microscopy measurements. Biophys J 2007;92:2926-43.

Cite This Article

Export citation file: BibTeX | RIS

OAE Style

Nolan JP, Chiu DT, Welsh JA. Rigor and reproducibility: status and challenges for single vesicle analysis. Extracell Vesicles Circ Nucleic Acids 2022;3:244-8. http://dx.doi.org/10.20517/evcna.2022.28

AMA Style

Nolan JP, Chiu DT, Welsh JA. Rigor and reproducibility: status and challenges for single vesicle analysis. Extracellular Vesicles and Circulating Nucleic Acids. 2022; 3(3): 244-8. http://dx.doi.org/10.20517/evcna.2022.28

Chicago/Turabian Style

Nolan, John P., Daniel T. Chiu, Joshua A. Welsh. 2022. "Rigor and reproducibility: status and challenges for single vesicle analysis" Extracellular Vesicles and Circulating Nucleic Acids. 3, no.3: 244-8. http://dx.doi.org/10.20517/evcna.2022.28

ACS Style

Nolan, JP.; Chiu DT.; Welsh JA. Rigor and reproducibility: status and challenges for single vesicle analysis. Extracell. Vesicles. Circ. Nucleic. Acids. 2022, 3, 244-8. http://dx.doi.org/10.20517/evcna.2022.28

About This Article

© The Author(s) 2022. Open Access This article is licensed under a Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, sharing, adaptation, distribution and reproduction in any medium or format, for any purpose, even commercially, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Data & Comments

Data

Views
556
Downloads
603
Citations
0
Comments
0
10

Comments

Comments must be written in English. Spam, offensive content, impersonation, and private information will not be permitted. If any comment is reported and identified as inappropriate content by OAE staff, the comment will be removed without notice. If you have any queries or need any help, please contact us at support@oaepublish.com.

0
Download PDF
Cite This Article 8 clicks
Like This Article 10 likes
Share This Article
Scan the QR code for reading!
See Updates
Contents
Figures
Related
Extracellular Vesicles and Circulating Nucleic Acids
ISSN 2767-6641 (Online)
Follow Us

Portico

All published articles are preserved here permanently:

https://www.portico.org/publishers/oae/

Portico

All published articles are preserved here permanently:

https://www.portico.org/publishers/oae/